MEDACTA INTERNATIONAL SA GMK-PRIMARY TIBIAL INSERT PS FIXED SIZE 3/10MM; KNEE TIBIAL INSERT Back to Search Results
Catalog Number 02.07.0310PSF
Health Effect - Clinical Code Pain (1994)
Date of Event 10/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clinical evaluation performed by medical affairs director: 4 years after primary total knee replacement the insert fixation screw got loose in the joint.Screw immediate removal and replacement is strongly recommended.No consequence is expected if no articular surface damage is visible.The cause for self-unscrewing of the insert screw is not known: one possibility is insufficient tightening torque, but other unknown conditions may also play a role.Batch review performed on 06 november 2018: lot 134298: (b)(4) items manufactured and released on 27-nov-2013.Expiration date: 2018-10-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
The patient came in complaining of pain 4 years and a half after primary.The surgeon determined that the screw backed out of the poly (right knee).The surgeon removed the screw.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-PRIMARY TIBIAL INSERT PS FIXED SIZE 3/10MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
MDR Report Key8041545
Report Number3005180920-2018-00858
Device Sequence Number1262834
Product Code JWH
UDI-Device Identifier07630030817069
UDI-Public07630030817069
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2018
Device Catalogue Number02.07.0310PSF
Device Lot Number134298
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2018
Initial Report FDA Received Date11/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-