BOSTON SCIENTIFIC CORPORATION PASSPORT; DILATOR, CATHETER, URETERAL Back to Search Results
Model Number M0062181100
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a passport balloon dilatation catheter was used during an upper ureteroscopy (urs) procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the balloon burst.The procedure was completed with another passport balloon dilatation catheter.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a passport balloon dilatation catheter was used during an upper ureteroscopy (urs) procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the balloon burst.The procedure was completed with another passport balloon dilatation catheter.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A microscopic examination identified a balloon longitudinal tear beginning approximately 20mm proximal of the balloon to distal tip bond and extending approximately 5mm proximately across the balloon material.An examination of the balloon material identified no issues which could potentially have contributed to this complaint.This failure indicates that an existing condition or disease is demonstrably responsible for the adverse event and use of the device has neither caused nor otherwise influenced this condition/disease-related adverse event.Therefore, the most probable root cause is adverse event related to patient condition.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the directions for use (dfu)/product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSPORT
Common Device Name
DILATOR, CATHETER, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8041594
Report Number3005099803-2018-61231
Device Sequence Number1262255
Product Code EZN
UDI-Device Identifier08714729191872
UDI-Public08714729191872
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K941853
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2020
Device Model NumberM0062181100
Device Catalogue Number218-110
Device Lot Number0021045840
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer11/12/2018
Initial Report FDA Received Date11/06/2018
Supplement Report FDA Received Date11/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
-
-