• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC RING TRANSJUGULAR INTRAHEPATIC LIVER ACCESS AND BIOPSY NEEDLE; GBZ CATHETER, CHOLANGIOGRAPHY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC RING TRANSJUGULAR INTRAHEPATIC LIVER ACCESS AND BIOPSY NEEDLE; GBZ CATHETER, CHOLANGIOGRAPHY Back to Search Results
Model Number N/A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #: exempt.Occupation: ir tech.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a ring transjugular intrahepatic liver access and biopsy needle was used during a transjugular intrahepatic portosystemic shunt procedure.When attempting to angle the needle into the target site, the needle would not turn (no difficulty with insertion per the customer).The device was removed from the patient and the needle tip separated into the sterile field.No portion of the device remained in the patient.The procedure was completed by using another needle.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Event or Problem Description
No new event description information to report at this time.
 
Additional Manufacturer Narrative
Investigation evaluation: a review of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, quality control, specifications, as well as a visual inspection and dimensional verification of the returned device was conducted during the investigation.One used rmt needle with a slight presence of biological matter was returned for evaluation.No surface damage was noted along the catheter or cannula.The proximal hub of the cannula was returned separated from the device and solder was present at the region of separation.The device passed manufacturing specifications that pertain to this failure, suggesting that there is no evidence the device was not manufactured to specifications.Additionally, a document based investigation evaluation was performed.It has been found that there are sufficient controls in place to detect the sheath tip failure mode prior to device release to the customer.A review of the device history record showed no relevant nonconformances that could have contributed to this failure mode.It should be noted that there were no other complaints reported for lot 9078493 or component lots.The instructions for use (ifu) advise the user to inspect the device prior to use to ensure that no damage has occurred.Based on the information provided, examination of the returned product and the results of our investigation, the root cause could possibly be related to the manufacturing of the device and a quality control deficiency.During the soldering process of the hub to the cannula, if a complete connection was not made, it could cause an inadequate seal that may lead to a separation if torsional force is applied.Also, if the user held the catheter too tight while trying to twist the needle, the needle would be difficult to twist.Adding then excessive torque to the hub to try and twist the needle could result in the solder joint breaking, causing the hub to separate.The appropriate personnel have been notified.Per the quality engineering risk assessment no further action is required.Cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RING TRANSJUGULAR INTRAHEPATIC LIVER ACCESS AND BIOPSY NEEDLE
Common Device Name
GBZ CATHETER, CHOLANGIOGRAPHY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8042768
Report Number1820334-2018-03315
Device Sequence Number1262876
Product Code GBZ
UDI-Device Identifier00827002314788
UDI-Public(01)00827002314788(17)210808(10)9078493
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2021
Device Model NumberN/A
Device Catalogue NumberRMT-16-50.5-RTPS-100-10.0
Device Lot Number9078493
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Initial Date Received by Manufacturer 10/23/2018
Supplement Date Received by Manufacturer12/21/2018
Initial Report FDA Received Date11/06/2018
Supplement Report FDA Received Date01/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-