Model Number LVZRMS180J
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the implant procedure, the guidewire unraveled when being removed from the left ventricular lead after it was placed into vein.The physician cut the wire at the proximal end of the lead, removed the lead and was able to extract the remaining portion of the wire from inside the lead.The lead was evaluated and determined to be undamaged and was re-implanted using a different guidewire.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the guidewire was returned in segments, and analyzed.Analysis indicated that the guidewire was unraveled.Visual analysis of the lead indicated damage during use.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZINGER GUIDE WIRE - CRDM
Common Device Name
OCCLUDER, CATHETER TIP
Manufacturer (Section D)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
MDR Report Key8043166
Report Number1220452-2018-00131
Device Sequence Number16488846
Product Code DQT
UDI-Device Identifier00643169452312
UDI-Public00643169452312
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K983927
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2021
Device Model NumberLVZRMS180J
Device Catalogue NumberLVZRMS180J
Device Lot NumberG18A08016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Initial Date Received by Manufacturer 10/16/2018
Supplement Date Received by Manufacturer03/01/2019
Initial Report FDA Received Date11/06/2018
Supplement Report FDA Received Date03/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
-
-