| Model Number |
MT22608-BLU |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, the receiver displayed "err hwbbt".No additional event or patient information is available.No product or data was provided for evaluation.The receiver has been received for device investigation.A follow-up report will be submitted upon completion of device investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.An external visual inspection was performed and passed.A receiver charge and boot test was performed and failed.The receiver data log could not be downloaded, but a "call tech support" error was observed and pictures were taken.The reported event of an error icon display was confirmed due to the occurrence of a "call tech support" error.A probable cause could not be determined.
|
| |
|