AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B2070-100
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The absolute pro stent and the other armada 35 balloon dilatation catheter referenced are filed under separate medwatch mfr numbers.Evaluation summary: visual and functional inspection was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other incidents.The investigation determined the reported balloon rupture appears to be related to circumstances of the procedure.It is likely that the balloon rupture was the result of interaction with anatomy which was described as heavily calcified.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during the mildly tortuous and heavily calcified left external iliac artery intervention, the 6.0 x 40 mm armada 35 balloon dilatation catheter (bdc) was advanced without resistance to the target lesion, but the balloon ruptured during the first inflation at 6 atmosphere (atm).A 7.0 x 100 mm armada 35 bdc was advanced without resistance to the target lesion, but at the first inflation the balloon ruptured at 7 atm.During deployment of the 8.0 x 100 absolute pro stent, the thumbwheel was difficult as it would stop intermittently, but the stent was successfully deployed.There was no reported adverse patient effect or a clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8043423
Report Number2024168-2018-08525
Device Sequence Number1302324
Product Code LIT
UDI-Device Identifier08717648155383
UDI-Public08717648155383
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Catalogue NumberB2070-100
Device Lot Number71207G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2018
Initial Report FDA Received Date11/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
-
-