Unknown taper.Medwatch sent to fda on (b)(6) 2018.The reporter of the event was asked to return the product for analysis.To date, apollo has not received the device.If returned, visual examination may confirm or determine another taper type associated with this event.Device labeling addresses the reported event as follows: precautions: it is the responsibility of the surgeon to advise the patient of the known risks and complications associated with the surgical procedure and implant.Adverse events it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Ulceration, gastritis, gastroesophageal reflux, heartburn, gas bloat, dysphagia, dehydration, constipation, and weight regain have been reported after gastric restriction procedures.Other adverse events considered related to the lap-band® system that occurred in fewer than 1% of subjects included: esophagitis, gastritis, hiatal hernia, pancreatitis, abdominal pain, hernia, incisional infection, infection, redundant skin, dehydration, gi perforation, diarrhea, abnormal stools, constipation, flatulence, dyspepsia, eructation, cardiospasm, hematemesis, asthenia, fever, chest pain, incision pain, contact dermatitis, abnormal healing, edema, paresthesia, dysmenorrhea, hypochromic anemia, band leak, cholecystitis, esophageal dysmotility, esophageal ulcer, esophagitis, port displacement, port site pain, spleen injury and wound infection.Warnings: patients should be advised that the lap-band ap® system is a long-term implant.Explant (removal) and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
|
|
Taper ii.Supplement #1: medwatch sent to fda on 12/28/2018.Device evaluation summary: a visual examination was performed on the returned lap-band with access port i and taper type ii.The band tubing was separated approximately 2.5" past the port tubing ss connector.The band tubing, port tubing, port taper, band ring and shell were noted to be discolored, as they were light brown in appearance.Needle marks were observed on the port septum and scratches were noted on one port hole.The band was separated in half approximately half way between the band buckle and band belt.The buckle strap was observed to be partially separated.An air leak test was performed, and leakage was noted from the band ring and shell where the band was separated in half.A fill inspection test was performed, and no blockage was noted when colored di water was passed through the port septum and tubing.Under microscopic analysis, both ends of the separated band ring and shell were noted to have a striated edges, consistent with damage from a surgical tool.Both ends of the partially separated portion of the buckle strap were observed to have striated edges, consistent with damage from a surgical tool.Both ends of the separated band tubing were observed to have striated edges, consistent with a surgical end cut to remove the device.Needle marks were noted on the port septum and on the port housing near the septum.Several non-penetrating nicks/marks were observed on one port hole.
|