Unknown taper.The reporter of the event was asked to return the product for analysis.To date, apollo has not received the device.If returned, visual examination may confirm or determine another taper type associated with this event.Device labeling addresses the reported event as follows: precautions: it is the responsibility of the surgeon to advise the patient of the known risks and complications associated with the surgical procedure and implant.Adverse events it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Ulceration, gastritis, gastroesophageal reflux, heartburn, gas bloat, dysphagia, dehydration, constipation, and weight regain have been reported after gastric restriction procedures.Gastroesophageal reflux, nausea and/or vomiting with early or minor slippage may be successfully resolved by band deflation in some cases.More serious slippages may require surgery to reposition and/or remove the band.Immediate re-operation to remove the band is indicated if there is total stoma-outlet obstruction that does not respond to band deflation or if there is abdominal pain.Other adverse events considered related to the lap-band® system that occurred in fewer than 1% of subjects included: esophagitis, gastritis, hiatal hernia, pancreatitis, abdominal pain, hernia, incisional infection, infection, redundant skin, dehydration, gi perforation, diarrhea, abnormal stools, constipation, flatulence, dyspepsia, eructation, cardiospasm, hematemesis, asthenia, fever, chest pain, incision pain, contact dermatitis, abnormal healing, edema, paresthesia, dysmenorrhea, hypochromic anemia, band leak, cholecystitis, esophageal dysmotility, esophageal ulcer, esophagitis, port displacement, port site pain, spleen injury and wound infection.Warnings: patients should be advised that the lap-band ap® system is a long-term implant.Explant (removal) and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
|
|
Taper ii.Supplement #1: medwatch sent to fda on 01/08/2019.Additional information: device evaluation summary: a visual examination was performed on the returned, lap-band with access port i, taper type ii.The band tubing still connected to the band was separated approximately 0.5 inches past the band collar.The band tubing connected to the port tubing was separated approximately 15" past the port tubing ss connecter.The band was noted to be separated near the location of the buckle.The band buckle was not returned with the device.Brown and white particles were observed on the inner surface of the band shell.White particles were noted on the outer surface of the band ring.Needle marks were observed on the port septum.See intake pic 5.Scratches were noted on the port housing.Wear and thinning was observed on the port tubing near the taper tubing junction.An air leak test was performed, and leakage was noted from the separated portion of the band near the location of the band buckle.A fill inspection test was performed, and no blockage was noted when colored di water was passed through the port septum and tubing.Under microscopic analysis, the end of the band tubing, separate approximately 0.6 inches past the collar, was noted to have striated edges, consistent with damage from a surgical end cut to remove the device.The end of the band tubing, separated approximately 15.9 inches past the ss connector, was noted to have striated edges, consistent with damage from a surgical end cut to remove the device.The end of the separated portion of the band was noted to have striated edges, consistent with damage from a surgical end cut to remove the device.Material degradation was noted on approximately 10% of the outer surface of the band.Wear and thinning was noted on a portion of the port tubing, approximately 0.5 inches in length near the taper tubing junction.Needle marks were noted on the port septum.Non-penetrating nicks/marks were noted on the port housing and the port base.
|