• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO CRESCENTIC BLADE (18.0 X 0.51 X 30.0MM); BLADE, SAW, GENERAL & PLASTIC SURGERY,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO CRESCENTIC BLADE (18.0 X 0.51 X 30.0MM); BLADE, SAW, GENERAL & PLASTIC SURGERY,SURGICAL Back to Search Results
Catalog Number 5400031417
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be filed once the quality investigation is complete.
 
Event or Problem Description
It was reported that there was a potential sterility breach on the packaging of the device.It was also reported that it was noticed prior to the procedure.It was further reported that there was no delay or adverse consequences as a result of this event.
 
Additional Manufacturer Narrative
Investigation results indicate that the reported failure of sterility breach (sticking through the packaging) was not confirmed.Markings were observed on the internal cavity surface of the blister where the teeth of the blade were in contact with it.It was visually clear that these markings were not a hole through the blister and only marking on the internal surface of the blister which could have been caused during handling and distribution.
 
Event or Problem Description
It was reported that there was a potential sterility breach on the packaging of the device.It was also reported that it was noticed prior to the procedure.It was further reported that there was no delay or adverse consequences as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRESCENTIC BLADE (18.0 X 0.51 X 30.0MM)
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY,SURGICAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key8044066
Report Number0001811755-2018-02373
Device Sequence Number11078232
Product Code GFA
UDI-Device Identifier04546540467256
UDI-Public04546540467256
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2020
Device Catalogue Number5400031417
Device Lot Number15267017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer12/13/2018
Initial Report FDA Received Date11/06/2018
Supplement Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-