• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC SYNFIX® EVO SPACER/MED DEEP 10.5MM HEIGHT/10°-STERILE; INTERVERTEBRAL FUSION DEVICE W/INTEGRATED FIXATION, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC SYNFIX® EVO SPACER/MED DEEP 10.5MM HEIGHT/10°-STERILE; INTERVERTEBRAL FUSION DEVICE W/INTEGRATED FIXATION, LUMBAR Back to Search Results
Model Number 08.815.241S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Expiration date is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.(b)(4).A review of the device history records has been requested.The investigation could not be completed; no conclusion could be drawn.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.
 
Event or Problem Description
T was reported that on (b)(6) 2018, anterior lumbar interbody fusion (alif) was performed at l4-s1 with synthes synfix evolution.There were difficulties placing one of the s1 screws on the l5-s1 level.After inserting the spacer at l5-s1, the awl was placed through the aiming device to cannulate s1.It was very difficult to fully advance the awl into the "schlerotic" bone that the patient had at s1.The awl appeared to be 15mm in length but the screws being used were 25mm.This left 10mm of bone not prepped for a screw.This was confirmed by viewing the ct scan on the imaging in the room - the bone was dense and schlerotic.After advancing the awl by malleting and twisting, a fine tip screw was placed into the hole of s1.After advancing the screw about 2/3 of the way through that plate on the spacer, the screw would not advance anymore.The screw was taken out.The surgeon wanted to drill into recannulated hole deeper into s1 so he could allow the screw to advance.With no drill option, the aiming device was elected to be taken off and the awl "freehand" was used which is not suggested in the technique guide.The screw was placed without the aiming device attached to the spacer.Using the standard screwdriver, the surgeon could only get the screw to advance about 2/3 of the way through the plate.Then, the straight screwdriver was used to get more axial pressure to advance the screw.And surgeon had to mallet occasionally to get the screw to advance, which ended up getting the screw to advance and lock into the plate.There were no issues with the remaining screws advancing using the awl through the aiming device.The procedure was completed successfully.There was a surgical delay of 10 minutes.Patient status is unknown.This complaint is for one (1) synfix® evo spacer/med deep 10.5mm height/10°-sterile.Concomitant device reported: unknown plate (part # unknown, lot # unknown, quantity 1).This complaint involves three (3) devices.This report is 1 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Dhr review was completed.Part: 08.815.241s.Lot: l227355.Manufacturing site: hägendorf.Release to warehouse date: 10.Jan.2017.Expiry date: 30.Nov.2026.The device history record shows this lot of 24 pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNFIX® EVO SPACER/MED DEEP 10.5MM HEIGHT/10°-STERILE
Common Device Name
INTERVERTEBRAL FUSION DEVICE W/INTEGRATED FIXATION, LUMBAR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8044109
Report Number2939274-2018-54873
Device Sequence Number19198410
Product Code OVD
UDI-Device Identifier10705034814285
UDI-Public(01)10705034814285
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K150673
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number08.815.241S
Device Catalogue Number08.815.241S
Device Lot NumberL227355
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/19/2018
Supplement Date Received by Manufacturer11/19/2018
Initial Report FDA Received Date11/06/2018
Supplement Report FDA Received Date11/19/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
-
-