• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRUETOME JAG 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRUETOME JAG 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00583140
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported that to boston scientific corporation that a truetome jag 44 was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the cutting wire broke.Reportedly, no part of the device detached inside the patient.The procedure was completed with a second truetome jag 44, using the same generator and active cord.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block a2: patient's exact age is unknown; however it was reported that the patient was over the age of 18.Block h6 (device codes): problem code 1069 captures the reportable event of cutting wire broken.Block h10: visual examination of the returned device revealed that the cutting wire was broken and blackened.There was no other issue noted.The complaint was consistent with the reported event of cutting wire broke.It is most likely that a peak of voltage could have caused the failures noted.Therefore, the most probable root cause is cause traced to component failure since the issue was associated with expected or random component failure without any design or manufacturing issue.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release to distribution.This supplemental correction report is being filed to correct the report number in a previously submitted supplemental report (follow-up number 001) which was accepted by fda on 12/18/2018 09:31 pm ct.The report number is being corrected from: 3005099803-2018-62043to: 3005099803-2018-61251.
 
Event or Problem Description
It was reported that to boston scientific corporation that a truetome jag 44 was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the cutting wire broke.Reportedly, no part of the device detached inside the patient.The procedure was completed with a second truetome jag 44, using the same generator and active cord.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETOME JAG 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8044606
Report Number3005099803-2018-61251
Device Sequence Number1421654
Product Code KNS
UDI-Device Identifier08714729858683
UDI-Public08714729858683
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K122203
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/20/2020
Device Model NumberM00583140
Device Catalogue Number47405
Device Lot Number0022146209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2018
Initial Date Received by Manufacturer 10/15/2018
Supplement Date Received by Manufacturer11/26/2018
Initial Report FDA Received Date11/06/2018
Supplement Report FDA Received Date01/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-