| Model Number |
9734715 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No device was returned to the manufacturer for analysis.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of procedure.It was reported that the tall spine clamp was loose, so the site changed it with a new one.There was no patient present when this issue was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis on the returned spinous process tall clamp resulted in no failure being found through functional testing and visual/physical examination.Analysis states that the clamp has no apparent physical damage and functions per design.If information is provided in the future, a supplemental report will be issued.
|
| |
|