MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTUA; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTUA; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V272
Device Problems Premature Discharge of Battery (1057); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/11/2018
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker declared elective replacement indicator less than two years after it was first implanted.Boston scientific technical services (ts) discussed the left ventricular lead exhibited an elevated output.As result, the battery consumption was faster than expected.Ts suggested replacing the device shortly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTUA
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CLONMEL LIMITED
cashel road
,
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
,
saint paul, MN 55112
6515826168
MDR Report Key8045349
MDR Text Key126356109
Report Number2124215-2018-61387
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526536663
UDI-Public00802526536663
Combination Product (y/n)N
PMA/PMN Number
P030005/S092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/22/2017
Device Model NumberV272
Device Catalogue NumberV272
Device Lot Number605100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/22/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age91 YR
-
-