| Model Number |
E SERIES |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The malfunction was observed during incoming testing; however after disassembling the device to determine root cause, the malfunction was no longer seen.The battery board to hv module cable was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device inappropriately shut down multiple times.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|