ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Low Audible Alarm (1016)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging quiet sounds issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the reported issue may cause the user to miss an alarm or warning that may cause cessation of insulin delivery.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 17-dec-2018 with the following findings: during investigation, the pump powered up with auditory and vibratory features.The alleged ¿quiet sound¿ issue was unable to be duplicated during testing.The pump cover was removed, an dno intermittent conditions were found to the piezo circuit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key8045682
Report Number2531779-2018-19225
Device Sequence Number16550491
Product Code OYC
UDI-Device Identifier10840406100082
UDI-Public10840406100082
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2018
Was the Report Sent to FDA? Yes
Device Age37 MO
Initial Date Received by Manufacturer 10/26/2018
Supplement Date Received by Manufacturer10/26/2018
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date12/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
-
-