MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX27522UX
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a moderately tortuous, severely calcified lesion exhibiting 70% stenosis located in the proximal ramus.The device was inspected with no issues.Negative prep was not performed.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device, however excessive force was not encountered during delivery of the device.It was reported that the device failed to cross the lesion, a guide liner was then used however the device still could not advance across the lesion.The device was removed from the patient.It was reported that stent deformation occurred in vivo during removal of the device from the patient.There was a stent strut lifted on the proximal segment of the stent.The patient is reported to be alive with no injury.
 
Additional Manufacturer Narrative
Product analysis summary: the stent was positioned between the marker bands but not meeting specifications due to deformation of the proximal stent segments.Deformation was evident to the most proximal stent wraps with struts raised.No deformation was evident to the distal tip.The inner lumen could not be verified with a 0.015 inch mandrel most likely due to hardened blood in the guidewire lumen.No other damage evident to the remainder of the device.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8045683
Report Number9612164-2018-03087
Device Sequence Number1787623
Product Code NIQ
UDI-Device Identifier00643169556911
UDI-Public00643169556911
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/09/2020
Device Catalogue NumberRONYX27522UX
Device Lot Number0009076638
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2018
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer12/06/2018
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date12/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-