• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN AVITUM ITALY S.P.A. APEX¿; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN AVITUM ITALY S.P.A. APEX¿; CONTAINER, I.V. Back to Search Results
Catalog Number 2112529
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Xemption number e2013017 b.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun italy s.P.A.(manufacturer).This report has been identified as event one of b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.Additional attempts to receive the sample are being made.A follow-up report will be submitted when the results of the investigation are available.
 
Event or Problem Description
As reported by user facility: event 1: found a very small white floater in a 500ml neonatal bag.No patient involvement.
 
Additional Manufacturer Narrative
Exemption number e2013017.B.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun italy s.P.A.(manufacturer).This report has been identified as event one of b.Braun medical inc.Internal report number (b)(4).Two (2) samples were received for evaluation.The bags were visually inspected and no evidence of any particles were found in either of the two bags.Afterwards, both samples were delivered to the biological laboratory for filtration on a black membrane, and no particles were found.Review of the device history record performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.Based on the results of the investigation, no conclusions can be made regarding the cause of the reported event.If additional information becomes available, a follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APEX¿
Common Device Name
CONTAINER, I.V.
Manufacturer (Section D)
B. BRAUN AVITUM ITALY S.P.A.
via xxv luglio 11
mirandola 41037
IT  41037
MDR Report Key8045984
Report Number9681240-2018-00020
Device Sequence Number1786959
Product Code KPE
UDI-Device Identifier04046964539121
UDI-Public(01)04046964539121
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K041415
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 12/18/2018
Report Date (Section F) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Catalogue Number2112529
Device Lot Number16G21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/18/2018
Device Age2 YR
Event Location Other
Date Report to Manufacturer12/18/2018
Initial Date Received by Manufacturer 10/18/2018
Supplement Date Received by Manufacturer10/18/2018
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date12/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA12/18/2018
-
-