Catalog Number 2112341
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Xemption number e2013017 b.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun italy s.P.A.(manufacturer).This report has been identified as b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.Additional attempts to receive the sample are being made.A follow-up report will be submitted when the results of the investigation are available.
 
Event or Problem Description
As reported by user facility: when getting prepared to spike the iv bag, the clinician noticed a gray/brown stain on the spike of the transfer set.Set was not used.
 
Additional Manufacturer Narrative
Exemption number e2013017 b.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun italy s.P.A.(manufacturer).This report has been identified as b.Braun medical inc.Internal report number 400408071.One (1) used set was returned for evaluation.On the pouch that the product was returned in it was written that dextrose was used in line 5.Under visual inspection traces of crystalline residue were present on the spike, and a brownish residue could be detected.There was not enough residue to be able to perform a ftir analysis.The tubing for line 5 also contained nutritional residue and other residues were present in line 5, indicating that the spike had already been inserted in a source container which could have been the source of the residue.A batch record review for the reported lot number did not reveal any abnormalities or non-conformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number.Based on the results of the investigation, no conclusions can be made regarding the cause of the reported event.If additional information becomes available, a follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE®
Common Device Name
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
B. BRAUN AVITUM ITALY S.P.A.
via xxv luglio 11
mirandola 41037
IT  41037
MDR Report Key8045994
Report Number9681240-2018-00019
Device Sequence Number19245950
Product Code LHI
UDI-Device Identifier04046964200229
UDI-Public(01)04046964200229
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K041222
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/21/2019
Report Date (Section F) 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Catalogue Number2112341
Device Lot Number18D09
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/23/2018
Device Age6 MO
Event Location Other
Date Report to Manufacturer01/21/2019
Initial Date Received by Manufacturer 10/23/2018
Supplement Date Received by Manufacturer10/23/2018
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date01/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/21/2019
-
-