• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number VPR-GW-FT14
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.(b)(4).
 
Event or Problem Description
Upon removal of the angioplasty balloon following treatment with an orbital atherectomy device (oad), it was discovered that the viperwire had fractured in more than one place.One fragment remained in the anterior tibial artery and two others remained in the dorsalis pedis artery.The fragments were removed with a combination of balloon and filter basket techniques following unsuccessful use of a snare.The patient was discharged home without complication following the procedure.
 
Additional Manufacturer Narrative
The guide wire was received at csi for analysis.Scanning electron microscopy analysis of the guide wire fractures revealed the presence of torsion dimples and possible fatigue shear dimples on the fracture face of the core wire faces.The core wire fracture sites also exhibited surface wear resulting in outside diameter reduction.The exact root cause could not be determined.The exact cause of the guide wire fracture is unknown.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
MDR Report Key8046185
Report Number3004742232-2018-00329
Device Sequence Number16540512
Product Code MCW
UDI-Device Identifier10852528005312
UDI-Public(01)10852528005312(17)200630(10)234377
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K151260
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberVPR-GW-FT14
Device Catalogue NumberVPR-GW-FT14
Device Lot Number234377
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer11/15/2018
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date12/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient Weight91
-
-