Model Number 88863324-56
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, while suturing the vulva, the thread on the first device broke and dropped.On the second device, the packaging was stuck and could not be removed.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of three images of one device.Photographic inspection noted that a suture was broken in two sites.Three suture ends were noted to be stretched and unbraided, a sign of splitting under tension.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI-CRON
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key8046295
Report Number9612501-2018-01951
Device Sequence Number1683210
Product Code GAT
UDI-Device Identifier20884521062549
UDI-Public20884521062549
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K930591
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model Number88863324-56
Device Catalogue Number88863324-56
Device Lot NumberD7D0286X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/12/2018
Supplement Date Received by Manufacturer04/11/2019
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date04/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-