| Model Number |
7120/75 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
Therapeutic Effects, Unexpected (2099)
|
| Date of Event |
07/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
During follow-up, increased impedance and noise leading to oversensing were observed on the device.The oversensing led to the delivery of inappropriate anti-tachycardia pacing (atp).No intervention was performed.The patient was in stable condition and there were no adverse consequences.
|
| |
|
Event or Problem Description
|
|
Additional information was received indicating that the right ventricular (rv) lead was explanted and replaced to resolve the event.
|
| |
|