| Catalog Number |
0279351250 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had continuous activation.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: cautery was live upon plugging into power unit, without pressing any buttons.The failure identified in the investigation is consistent with the complaint record.The probable root cause/s could be the detachment of the suction tube to the suction shaft causing a leak internally, which creates a short and a possibility would result in continuous activation.Inadequate gluing operation on the suction tube and suction adapter.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had continuous activation.
|
| |
|