It was reported, a physician was planning to insert a stent graft into the patient by approaching from the left femoral artery.When he was shaping the distal end of the amplatz extra-stiff support heparin coated wire guide, it became elongated.This device never made patient contact.The procedure was completed with another amplatz extra-stiff support heparin coated wire guide to.There have been no adverse effects to the patient reported.According to the complainant, the complaint device was discarded by the user facility.It is therefore, not available to assist the manufacturer with the investigation into this event.
|
|
Blank fields on this form indicate the information is unknown, or unavailable.Investigation evaluation.A review of the complaint history, device history and documentation, drawings, manufacturing instructions, specifications and quality control were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.The device history record was reviewed, two non-conformances were noted for flexibility incorrect (qty 2), and relaxation incorrect (qty 23).While these nonconformances are related to the reported failure mode, the affected units were scrapped and not replaced.A complaint history search revealed one other complaint associated to the lot number 9009107 for abnormal tip shape.Furthermore, a review of the manufactures instructions and quality control procedures was conducted, and no gaps were discovered.Based on information reported from facility , prior to patient contact the physician attempted to shape the distal end of the wire guide and it elongated.The investigation conclusion assessed the physician's shaping of the tip as a potential cause of tip elongation as the failure mode occurred after manipulation.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|