| Catalog Number |
0580-1-443 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
|
| |
|
Event or Problem Description
|
|
Surgeon was placing definitive exeter implant onto the stem introducer.I noticed he had placed the winged centraliser onto the stem for a small canal.We discussed his option to use the straight centraliser to use with alongside the small artisan plug.He acknowledged my suggestion but chose to go ahead with the winged centraliser.He then attached the implant onto the stem introducer and wrapped a pack over the end of the stem and placed it on the table still wrapped in the pack.He then picked up the stem, wrapped in the pack so that the tip was not visible and proceeded to insert the stem into the patient/cement.It was then noticed by the stryker representative that the centraliser was on the table behind some retractors.The rep brought this to the immediate attention of the surgeon to which he replied, ¿oh well, too late now¿.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding off label use involving an exeter stem was reported.Conclusion: a review of the exeter® v40® femoral stem using exeter broach, exeter-(b)(4) noted the following: the hollow centraliser must be used with the exeter stem because it provides a space below the stem tip, which prevents ¿end-bearing¿ of the stem and ensures that the proximal, expanded taper of the stem will engage properly in the cement mantle.Each exeter v40 stem is supplied with a winged and a straight-sided centraliser.Based on the provided information it has been determined that this event is associated with an off-label application.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Surgeon was placing definitive exeter implant onto the stem introducer.I noticed he had placed the winged centraliser onto the stem for a small canal.We discussed his option to use the straight centraliser to use with alongside the small artisan plug.He acknowledged my suggestion but chose to go ahead with the winged centraliser.He then attached the implant onto the stem introducer and wrapped a pack over the end of the stem and placed it on the table still wrapped in the pack.He then picked up the stem, wrapped in the pack so that the tip was not visible and proceeded to insert the stem into the patient/cement.It was then noticed by the stryker representative that the centraliser was on the table behind some retractors.The rep brought this to the immediate attention of the surgeon to which he replied, ¿oh well, too late now¿.
|
| |
|