• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FERNO-WASHINGTON, INC. INX FOR INLINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FERNO-WASHINGTON, INC. INX FOR INLINE Back to Search Results
Model Number 0015803
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/25/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported while loading the empty cot into the ambulance the operator end legs were raised; however, the medic operator thought the head end legs had raised and upon pushing the cot into the truck the cot allegedly disengaged from the safety hook and the operator end of the cot suddenly lowered and pinned the medic's wrist under the cot handle.The medic alleged a wrist and hand injury and was seen in the emergency room.The incident report alleges a suspected fracture and thumb strain.
 
Event or Problem Description
It was reported while loading the empty cot into the ambulance the operator end legs were raised; however, the medic operator thought the head end legs had raised and upon pushing the cot into the truck the cot allegedly disengaged from the safety hook and the operator end of the cot suddenly lowered and pinned the medic's wrist under the cot handle.The medic alleged a wrist and hand injury and was seen in the emergency room.The incident report alleges a suspected fracture and thumb strain.
 
Additional Manufacturer Narrative
An authorized field technician was dispatched to the complainant's location to evaluate the cot.A visual and functional evaluation was conducted and the cot was found to be operating as intended.There were no observations of damage or defects.A function test was also conducted by the manufacturer on a similar device and the reported incident could be duplicated by improperly loading the cot by lifting the foot end of the cot and pushing the raise button before pushing the cot into the ambulance.This would cause the operator end legs to raise and the operator would have to assume the weight of the cot.Proper loading instructions are contained in the ifu as well as multiple warnings against assuming the weight of the cot prior to pushing the head end of the cot into the ambulance.The manufacturer has scheduled additional training to be conducted with the service personnel to ensure proper loading technique.No additional information was provided regarding the medic's alleged injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INX FOR INLINE
Common Device Name
INX FOR INLINE
Manufacturer (Section D)
FERNO-WASHINGTON, INC.
70 weil way
wilmington OH 45177
MDR Report Key8048402
Report Number1523574-2018-00035
Device Sequence Number1683875
Product Code FPO
UDI-Device Identifier00190790000004
UDI-Public00190790000004
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number0015803
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/30/2018
Supplement Date Received by Manufacturer10/30/2018
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-