• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT HT VERSATURN GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT HT VERSATURN GUIDE WIRE Back to Search Results
Catalog Number 1013317J
Medical Device Problem Codes Entrapment of Device (1212); Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned for analysis.The reported detachment was confirmed.The reported entrapment was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the deployed stent resulted in the reported entrapment.Manipulation of the device resulted in the reported detachment.As reported, the versaturn fragment remains under the implanted stent.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during a left anterior descending (lad) artery intervention, the versaturn guide wire was positioned in the diagonal branch as a vessel stabilizer.A non-abbott stent was implanted and post-dilatation was completed.During removal of the versaturn without resistance, the guide wire separated distally and was entrapped under the implanted stent and slightly protruding into the diagonal branch.Another non-abbott stent was implanted to stabilize the detached portion of the versaturn wire.The patient was reported as ok and the outcome reported as normal.The patient was discharged the next day.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HT VERSATURN GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8048496
Report Number2024168-2018-08571
Device Sequence Number1625155
Product Code DQX
UDI-Device Identifier08717648198472
UDI-Public08717648198472
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K103101
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number1013317J
Device Lot Number8073061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2018
Initial Report FDA Received Date11/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient Weight86
-
-