JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZKB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the multifocal zkb00 +13.0 diopter intraocular lens (iol) was implanted into the patient's right eye on (b)(6) 2018.Reportedly, the patient was experiencing dysphotopsia therefore the lens was explanted on (b)(6) 2018 and replaced with a monofocal pcb00 +13.5 diopter lens.No incision enlargement, no vitrectomy was performed, no sutures were used and no patient injury was reported.The patient was reported in good spirit when discharged.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8048582
Report Number9614546-2018-01058
Device Sequence Number1654572
Product Code MFK
UDI-Device Identifier05050474559943
UDI-Public(01)05050474559943(17)190324
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/24/2019
Device Model NumberZKB00
Device Catalogue NumberZKB00U0130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/12/2018
Supplement Date Received by Manufacturer11/14/2018
10/25/2020
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date11/22/2018
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Date Report Sent to FDA01/01/2005
-
-