It was reported that the patient had a heart rate of 37 beats per minute (bpm), when the lower rate limit was programmed at 40 bpm.Technical services discussed possible causes.The healthcare provider health care professional (hcp) was going to bring the patient in, to learn all the parameters.Additional information was reported, indicating that there was a loss of capture (loc) at 5 volts (v), on the right ventricular (rv) epicardial lead from another manufacturer.However, current threshold was 0.8 v.The field representative stated that the clinic programmed the lead configuration to unipolar pace and bipolar sense, and left the output at 5v at 0.4 milliseconds.There was an occasional loc still observed.There were also reportedly some random pace impedance spikes, thus the minute ventilation (mv)feature was programmed to off.Data analysis revealed that the pace impedance spikes were within normal limits and there were no device issues.Further review of the previous electrograms (egm) revealed noise oversensing as a result of the mv feature.At one point, the patient had reportedly experienced a 14 second pause.This occurred prior to the device reprogramming.No additional adverse patient effects were reported.This device was included in the recent minute ventilation sensor signal oversensing advisory population.
|