• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FUSE; ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FUSE; ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT Back to Search Results
Model Number FSC330-F7526
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although expected, the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a fuse c38 colonoscope - s was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the scope presented a hazy center image.The patients anatomy was not visible on the center screen.The procedure was completed with another fuse c38 colonoscope - s.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a fuse c38 colonoscope - s was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the scope presented a hazy center image.The patients anatomy was not visible on the center screen.The procedure was completed with another fuse c38 colonoscope - s.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Device code 1304 captures the reportable event of center image lost.A fuse c38 colonoscope was received for analysis.A functional evaluaton was performed and moisture was found in the forward objective lens due to glue deterioration around the seal.The reported issue was confirmed.The cause of reported issue is due to moisture found.A damage seal could contributed to the reported event.Therefore, the assigned investigation conclusion code for this complaint is designated as cause traced to component failure.The repair history was reviewed and nothing was found to indicate a possible service-related cause for the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSE
Common Device Name
ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8048706
Report Number3005099803-2018-61262
Device Sequence Number16503404
Product Code PEA
UDI-Device Identifier00840253109552
UDI-Public00840253109552
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K160275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFSC330-F7526
Device Catalogue Number60093
Device Lot Number2300252
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2018
Initial Date Received by Manufacturer 10/11/2018
Supplement Date Received by Manufacturer12/13/2018
Initial Report FDA Received Date11/07/2018
Supplement Report FDA Received Date01/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-