| Model Number |
MMT-523RNAH |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform the displacement test due to completely broken reservoir tube lip.Unit had missing reservoir tube o-ring.
|
| |
|
Event or Problem Description
|
|
Customer's son reported via phone call that there was damage on reservoir lip.The customer¿s blood glucose level was 175 mg/dl.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|