| Catalog Number |
RONYX27508UX |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
10/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a severely tortuous and calcified lesion exhibiting 90% stenosis located in the ostium circumflex (cx) artery.The device was inspected with no issues.Negative prep was performed with no issues.The lesion was pre-dilated.The device passed through a previously deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery of the device.It was reported that stent dislodgement occurred during the removal of the device following a failed delivery.The dislodged stent was removed from the patient using a snare.The patient is reported to be alive with no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The inflation device remained on neutral pressure during delivery of the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|