| Model Number |
E2781R-28ASP |
| Medical Device Problem Codes |
Melted (1385); Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic colectomy, there was no activation from the tip of the device, although there was activation of the tissue from the black insulated part which touched the tissue which the insulation was melted near the tip of the shaft.Nothing fell into the patient's cavity.The procedure was completed with another similar device.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation.The returned product did not meet specification as received.Visual inspection found the device to have melted near and above the aspiration holes.The reported condition was confirmed.The investigation found a section of the insulation around the irrigation holes to have melted.No insulation appears to be missing from the device.The device failed hipot testing around the damaged insulation indicating electrical leakage.The melting occurred as a result of arcing.Two possible sources of arcing were identified.Simultaneously activating the suction function and the electrical current may result in increased arcing at either the electrode tip or aspiration holes, causing the insulation to melt.Activating the electrode while retracted in the sheath could result in arcing through the aspiration holes and potential melting.The investigation identified the root cause of the reported event to be attributed to the user.The instructions for use (ifu) states, simultaneously activating suction/irrigation and electrosurgical current may result in increased arcing at either the electrode tip or aspiration holes, and/or burns to adjacent tissue.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|