US SURGICAL PUERTO RICO ROTICULATOR; STAPLE, IMPLANTABLE Back to Search Results
Model Number 017614
Medical Device Problem Codes Misfire (2532); Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during an open procedure, the device was fired, then stopped in the distal area after the first shot.They opened another new stapler to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.There were no visual or functional abnormalities noted during pmv testing.The instrument was received with the pushers advanced and the handle locked.Functionally, the instrument was reset and cycled.The jaw properly clamped and unclamped when the approximating lever was operated.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTICULATOR
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key8050955
Report Number2647580-2018-05408
Device Sequence Number19243636
Product Code GDW
UDI-Device Identifier10884521047709
UDI-Public10884521047709
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K855047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Model Number017614
Device Catalogue Number017614
Device Lot NumberP6E0626X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2019
Initial Date Received by Manufacturer 10/15/2018
Supplement Date Received by Manufacturer04/25/2019
Initial Report FDA Received Date11/08/2018
Supplement Report FDA Received Date05/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-