| Model Number |
017614 |
| Medical Device Problem Codes |
Misfire (2532); Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open procedure, the device was fired, then stopped in the distal area after the first shot.They opened another new stapler to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.There were no visual or functional abnormalities noted during pmv testing.The instrument was received with the pushers advanced and the handle locked.Functionally, the instrument was reset and cycled.The jaw properly clamped and unclamped when the approximating lever was operated.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|