• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD Back to Search Results
Model Number 7070M-90
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 10/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is user error; improper implant size selection, using an implant that was too long or placing the implant too deep.
 
Event or Problem Description
The patient had left side si joint arthrodesis in (b)(6) 2017 where two implants were placed.The patient complained of radicular pain following the procedure.X-rays indicated that the most cranial positioned implant was impinging on the neuroforamen.Two days after the initial procedure, the surgeon performed a revision procedure where he removed the cranial positioned implant and added two new implants in different positions.No other preexisting implants were adjusted or removed.The status of the patient following the revision procedure is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
MDR Report Key8051334
Report Number3007700286-2018-00101
Device Sequence Number11159579
Product Code OUR
UDI-Device Identifier00851085007541
UDI-Public00851085007541
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/11/2022
Device Model Number7070M-90
Device Lot Number2585602
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2018
Initial Report FDA Received Date11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
-
-