| Catalog Number |
283709 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If additional information should become available, a supplemental medwatch will be submitted accordingly.(b)(4) - incomplete.The exp date is currently unavailable.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
This is report 2 of 2 for the same event.It was reported by the affiliate in (b)(6) that during an unspecified surgical procedure of the shoulder repair, it was observed that two full radius 5.0mm 5pk devices had metal abrasion.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown but was noted to have occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Associated medwatch report number: 1221934-2018-55238.Corrected data: date of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and evaluated.Visual observation of device revealed the presence of the white lubricant and blades were found rusted.The white creamy substance detected is teflon based lubricant and is intended to be on the shafts of the blades to reduce friction between the inner and outer shaft and to mitigate failures such as inner and outer blades sticking to each other and metal shavings.The inner and outer sleeves were separated and inspected for any damage.The devices had signs of friction and striation marks on the surface of the distal end of the inner sleeve.The friction marks indicate friction between the sleeves, which could lead to metal shavings as reported.Therefore, the reported failure can be confirmed.One possible root cause for the reported condition could be the excessive side loading during the use.Other than above mentioned possibility, we cannot discern any definite root cause at this point of time.A white paper was performed previously on this failure; the result indicated that the amount of shedding for this product is acceptable.Review of the device history record indicated that this batch of product was processed without incident; therefore there is no evidence of manufacturing anomalies on the records reviewed.At this time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi:(b)(4).Associated medwatch report number: 1221934-2018-55238.
|
| |
|