An end user alleged he developed a "foot infection" while using a dreamwear nasal mask, dreamstation auto continuous positive airway pressure (cpap) device, and associated humidifier.The user stated the materials in the mask caused redness to his face, and subsequently he developed an infection in his foot after using the devices for one week.The user reported being given antibiotics and has since recovered.The manufacturer's investigation is ongoing.Upon completion of the investigation, a follow up report will be filed.
|
|
The manufacturer received the dreamwear nasal mask, cpap, and associated humidifier for investigation.There were no visual defects noted to the dreamwear mask, and the cpap and humidifier functioned as designed.There were no obvious contaminants that could have led to the user's reported "foot infection".The certificate of conformance for the mask indicates it was manufactured to meet all specifications and standards.This mask meets all relevant standards for biocompatibility.A review of the data logged by the cpap device indicates it was in use for a period of only three days.The dreamwear nasal mask is intended to provide an interface for application of cpap or bi-level therapy to patients.The mask is for single patient use in the home or multi-patient use in the hospital/institutional environment.The mask is to be used on patients (>(b))(4)/) for whom cpap or bi-level therapy has been prescribed.The user is warned to discontinue use and contact a health professional if skin redness, irritation, or discomfort occur.The user is also instructed to wash the mask prior to use.The philips respironics dreamstation systems deliver positive airway pressure therapy for the treatment of obstructive sleep apnea in spontaneously breathing patients weighing over (b)(4) it is for use in the home or hospital/institutional environment.Labeling for the cpap and humidifier instructs the user, "if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, if it has been dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, disconnect the power cord and discontinue use.Contact your home care provider." there were no operational issues noted with the dreamwear mask, cpap, or humidifier.The manufacturer is unable to confirm any malfunction that may have led to the user's allegation that he developed a "foot infection" while using these devices for three days.Based on the information available, the manufacturer concludes no further action is necessary.
|