| Model Number |
15 |
| Medical Device Problem Code |
Inappropriate or Unexpected Reset (2959)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control performed an initial evaluation of the customer's device and verified the reported issue.The customer will receive a replacement device.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device re-boots itself approximately every 30 seconds and will not display ecg.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control further evaluated the customer's device and determined the cause for the reported issue was due to the system pcb assembly.The customer received a replacement device.Physio-control further evaluated the removed system pcb assembly and verified the reported issue.It was determined that the system pcb assembly caused the device to continually re-boot itself, ecg could not be displayed and the device could not charge and shock; however further cause could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device re-boots itself approximately every 30 seconds and will not display ecg.There was no patient use associated with the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device re-boots itself approximately every 30 seconds and will not display ecg.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Section e4 initial report sent to fda of the initial medwatch report indicates: blank section e4 initial report sent to fda of the initial medwatch report should indicate: no.
|
| |
|