Catalog Number 0066204
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No errors were noted in event log for the echo v2 instrument.Based off results obtained and review of all reacting cells on all panels tested.Sample appears to be showing dosage for big e.However, unable to rule out compromised pouch of capture-r ready-id lot number id368.No sample or product was returned to immucor for evaluation.The immucor internal record number for this report is (b)(4).
 
Event or Problem Description
On (b)(6) 2018 a customer reported unexpected negative reactivity with capture-r ready-id on the echo v2 instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY-ID
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key8052501
Report Number1034569-2018-00254
Device Sequence Number16555267
Product Code KSZ
UDI-Device Identifier10888234001140
UDI-Public10888234001140
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
102707/0.0
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date12/25/2018
Device Catalogue Number0066204
Device Lot NumberID368
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2018
Initial Report FDA Received Date11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
-
-