• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 382554
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported with the use of the bd insyte¿ autoguard¿ bc shielded iv catheter there was an issue with sterile breach.There was no report of injury or medical intervention.
 
Additional Manufacturer Narrative
H.6.Investigation summary: dhr- the product was manufactured on afa line 9 from march 24, 2017 thru march 25, 2017 packaged on packaging line 11 from april 5, 2017 thru april 6, 2017.All challenge samples, set-up and in process testing were performed and all passed per specifications.No quality notifications were initiated during production.Received one iag unit from lot number 7076873, all components within were present and intact.The unit received was partially open at one end.The product characteristics require a minimum of 1/8¿ seal width with adhesive transfer from the top web paper to the bottom web film.This characteristic was met.The key variables that affect the packaging seal are seal transfer/width and top web adhesive presence.Both of these variables were included in the investigation.Conclusion(s): supplier ¿ defective material bd supplier oliver-tolas (ot) 29lp uses a standard reinforced paper that is common to many other suppliers, but the adhesive type and application is specific to oliver-tolas.There is sufficient evidence to demonstrate the ot material or adhesive application is the root cause.Due to continuous issues with ot, bd is moving to an alternate supplier for the top web material.Capa 48637 was opened to investigate the package seal integrity complaints and implement corrective actions.
 
Event or Problem Description
It was reported with the use of the bd insyte¿ autoguard¿ bc shielded iv catheter there was an issue with sterile breach.There was no report of injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
MDR Report Key8052547
Report Number1710034-2018-00809
Device Sequence Number19322517
Product Code FOZ
UDI-Device Identifier30382903825548
UDI-Public30382903825548
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K110443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/29/2020
Device Catalogue Number382554
Device Lot Number7076873
Initial Date Received by Manufacturer 10/22/2018
Supplement Date Received by Manufacturer10/22/2018
Initial Report FDA Received Date11/08/2018
Supplement Report FDA Received Date01/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-