• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM STANDARD; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM STANDARD; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Catalog Number PRB35-05-080-120
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the physician intended to use a protégé everflex self expanding stent to treat a cto in the right superficial femoral artery(sfa).The lesion was crossed with a wire and atherectomy performed followed by balloon inflation.The physician then inserted the self expanding stent.The stent was advanced and deployed successfully.It was reported the delivery catheter was withdrawn, a device component was observed by angiogram at the lateral branch of the profunda artery proximal to the deployed stent.Attempts to snare the component was unsuccessful.Angiogram shows foreign body in the laterial branch of the profunda artery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM STANDARD
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
MDR Report Key8054563
Report Number2183870-2018-00508
Device Sequence Number1653679
Product Code FGE
UDI-Device Identifier00821684068588
UDI-Public00821684068588
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K060057
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/18/2020
Device Catalogue NumberPRB35-05-080-120
Device Lot NumberA526244
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2018
Initial Report FDA Received Date11/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-