• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 44MM NO 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 44MM NO 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 0580-1-443
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
 
Event or Problem Description
A fractured stem neck.
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving an exeter stem was reported.The event was confirmed through review of an x-ray by a clinician.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: a review of the provided medical records and x-rays by a clinical consultant indicated: x-ray confirms fractured neck, need operative reports.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
A fractured stem neck.The patient was standing up doing some households when it suddenly snapped off.Along with this came pain and the patient lost balance.The revision took place (b)(6).Another exeter 50 offset 125 length was put in together.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXETER V40 STEM 44MM NO 3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8054822
Report Number0002249697-2018-03668
Device Sequence Number11131062
Product Code JDI
UDI-Device Identifier04546540153326
UDI-Public04546540153326
Combination Product (Y/N)N
Initial Reporter CountrySE
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Catalogue Number0580-1-443
Device Lot NumberG3753762
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/12/2018
Supplement Date Received by Manufacturer02/18/2019
Initial Report FDA Received Date11/09/2018
Supplement Report FDA Received Date03/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-