| Model Number |
1884004HR |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
Radiation Exposure, Unintended (3164); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis found that visually, the distal tip had broken off the inner cutter.The portion that became detached was not returned however it would have measured approximately 0.24¿ in length.The break occurred at the first proximal tooth valley.The outer assembly was deformed.There was wear around the circumference of the inner shaft 0.55¿ from the distal face of the inner hub.There was uneven wear at the cutting tip.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported via a manufacturer representative that a small piece of blade broke off inside the sheath into the patient during a submucous resection procedure.A new blade was opened and the case was completed.There was delay in the procedure.The missing piece was not located after several xrays.The patient is fine and had no issues after the procedure.
|
| |
|
Search Alerts/Recalls
|