A customer from (b)(6) reported the generation of an invalid cobas mpx positive control [mpx o (+) control] due to an error flag (c02h1).When reviewing several runs within the customer-provided data, it was identified that several wells, containing samples or controls, had the error code p07p, which is indicative of a volume error during supernatant removal.The roche field service engineer visited the customer site and found a failed result for process head tightness check with some loose stop discs on the cobas 6800 system's processing transfer heads, that were likely the cause for the generation of the error codes.The roche field service engineer replaced all stop discs & o-rings on this instrument processing transfer head.There is no allegation on discrepant donor sample results.When error codes are generated for the cobas mpx controls, the test run is invalidated and all donor samples included in the test run need to be repeat tested per the instructions for use.When the p0p7 error code is generated for a donor sample, the donor sample needs to be repeat tested per the instructions for use.No harm or injury was reported through the case.
|