It was reported that a (b)(6)-year-old male patient, underwent a right knee arthroscopy procedure with a shaver blade full-radius resector large hub yellow and second-degree burn occurred.The patient was brought to the operating room for a knee scope on the right knee.Procedure was going as planned with no issues and surgeon was using a full radius resector when both the surgical technician and the physician assistant (pa) noticed a burning smell.Scrub looked down and saw some smoke coming from the resector and immediately noticed that it was broken, no fire was noticed.She notified the surgeon, who stopped what he was doing and removed the instrument.The resector was inspected and found to be obviously broken.Surgeon and pa started to inspect the knee at the portal site and noted that there was some possible tearing.It was later determined that it was most likely a second degree burn at the incision site where the shaver was being used in the knee.The spot was measured, and pictures were taken.A new device, 3.5 fill radius shaver was opened, and the case continued and completed successfully without further incident.The physician¿s opinion on the origin of the smoke it was caused by heat generation and the burn was caused by the shaver.There was no specific procedural or patient circumstance that contributed to the break of the device.The patient was seen and treated at the facility wound clinic, however exact treatment is unknown.
|
|
It was reported that a (b)(6) male patient, underwent a right knee arthroscopy procedure with a shaver blade full-radius resector large hub yellow and second-degree burn occurred.The product said to be involved was not received for evaluation.As the device was not returned for evaluation a thorough examination could not be performed.The device history record was reviewed and the records show that the device passed all visual & functional testing prior to being shipped to the account.A photograph of the involved product was provided.The image is too low a resolution (as well as out of focus) to zoom in on sections of the device to conduct an effective analysis.However, it can be assessed that the device's shaft, with the inner cannula fully inserted within the outer cannula, is bent to a significant degree.This is indicative of excessive force on the device's structure, likely from excessive torqueing of the product while in use.Unable to evaluate the bend to determine if this is a full bend or a partial/complete break and involves the butt weld joint (where two pieces of the cannula are welded together on the same plane) a characteristic of linvatec shavers.While the area appears bent, no conclusion could be made as to whether the device made contact with another metal device or hard object.The account reported that the device ¿burnt the patient¿.Cannot determine from the photo, how heat would have affected or came from the product.As the device was used on the patient, no conclusion could be determined as to a possible cause for their reported failures.The instructions for use for reprocessed powered arthroscopic accessories, which provides various warnings and precautions, it is noted to state, "do not operate the cutter, shaver, or burr at rpms in excess of those recommended.Do not operate without the flow of irrigation.Periodic irrigation of the tip of the burr or cutter is recommended to provide adequate cooling of the blade and to prevent accumulation of excised materials in the surgical site.Do not contact cutters, shavers, or burrs with metal objects as this may cause the device to break or shed metal.Excessive ¿side-loading¿ on the blade during use does not improve cutting performance, and in extreme cases, may result in wear and degradation of the inner blade.Excessive pressure may cause cutters, shavers, or burrs to bend or break which may in turn cause harm to the patient and/or operating room staff.¿ if device is received, a complete evaluation will be conducted.Manufacturer's ref.No: (b)(4).
|