• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. TRIVEX SYSTEM RESECTOR HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. TRIVEX SYSTEM RESECTOR HANDPIECE Back to Search Results
Catalog Number 7210387
Medical Device Problem Code Failure to Conduct (1114)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have received the device for evaluation.The housing and power cord were visually inspected and were found to be acceptable.The handpiece status light flashed green when the device was connected to the control unit indicating an issue with the handpiece.The drive shaft did not rotate when the window lock button was pressed.However, the drive shaft rotated properly when the run button was pressed.Upon disassembly of the housing, water was observed inside the core tube.The window lock reed switch on the switch pc board was also found to be malfunctioning.The root cause of the defect was determined to be a damage to the pc board due to water exposure.We currently have a corrective and preventive action (capa) open to address this issue.There was no injury to the patient.The malfunction was detected prior to the case.
 
Event or Problem Description
During pre-use check, the resector on the handpiece did not rotate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIVEX SYSTEM RESECTOR HANDPIECE
Common Device Name
HANDPIECE
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 02148
MDR Report Key8056680
Report Number1220948-2018-00098
Device Sequence Number2190113
Product Code DWQ
UDI-Device Identifier00840663107391
UDI-Public00840663107391
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K032387
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number7210387
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2018
Initial Report FDA Received Date11/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-