• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK EXPRESS(R) DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK EXPRESS(R) DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number EXPRESS
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control performed an initial evaluation of the customer's device and verified the reported issue.The customer will be provided with a replacement device.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted physio-control to report that their device had a service wrench icon present on its readiness display.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that all three icons (charge pak, attention and service wrench) were present on its readiness display and that it would not power on when prompted.
 
Event or Problem Description
The customer contacted physio-control to report that their device had a service wrench icon present on its readiness display.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that all three icons (charge pak, attention and service wrench) were present on its readiness display and that it would not power on when prompted.
 
Additional Manufacturer Narrative
Physio-control further evaluated the customer's device and determined that the cause of the reported issue was a shorted capacitor, designator c76, on the digital pcb assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK EXPRESS(R) DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key8057207
Report Number0003015876-2018-01877
Device Sequence Number1915455
Product Code MKJ
UDI-Device Identifier00885074734453
UDI-Public00885074734453
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K033275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEXPRESS
Device Catalogue Number99427-000129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2018
Initial Date Received by Manufacturer 10/16/2018
Supplement Date Received by Manufacturer12/03/2018
Initial Report FDA Received Date11/09/2018
Supplement Report FDA Received Date12/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-