| Model Number |
EXPRESS |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control performed an initial evaluation of the customer's device and verified the reported issue.The customer will be provided with a replacement device.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had a service wrench icon present on its readiness display.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that all three icons (charge pak, attention and service wrench) were present on its readiness display and that it would not power on when prompted.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had a service wrench icon present on its readiness display.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that all three icons (charge pak, attention and service wrench) were present on its readiness display and that it would not power on when prompted.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control further evaluated the customer's device and determined that the cause of the reported issue was a shorted capacitor, designator c76, on the digital pcb assembly.
|
| |
|