Model Number ZXT225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported zxt225 23.0 diopter intraocular lens (iol) was implanted in the patient¿s ocular sinister (left eye) on (b)(6) 2018 and explanted on (b)(6) 2018 due to patient complaint of unknown/unspecified visual issues.Through follow up, customer account provided additional information confirming no incision enlargement, no serious patient injury, no suture(s), no vitrectomy, and the replacement lens was a zxr00 22.0 diopter iol.Patient outcome was reported 20/30 visual acuity.No additional information was provided to johnson & johnson surgical vision.
 
Additional Manufacturer Narrative
Product testing could not be performed because the product was not returned.The customer's reported event could not be confirmed.During the manufacturing record review of the production order for this serial number, no product deficiency was identified.The documentation shows that the production order was manufactured according to specifications.A search of complaints revealed no similar complaints were received for this production order number.As a result of the investigation there is no indication of a product malfunction or product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION
1700 e. st. andrew place
santa ana CA 92705
MDR Report Key8057310
Report Number9614546-2018-01065
Device Sequence Number1921028
Product Code POE
UDI-Device Identifier05050474605855
UDI-Public(01)05050474605855(17)230420
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/20/2023
Device Model NumberZXT225
Device Catalogue NumberZXT225U230
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/15/2018
Supplement Date Received by Manufacturer11/15/2018
10/25/2020
Initial Report FDA Received Date11/09/2018
Supplement Report FDA Received Date12/12/2018
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Date Report Sent to FDA01/01/2005
-
-