Fracture (1260); Material Twisted/Bent (2981); Device Handling Problem (3265)
Health Effect - Clinical Code
No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event
Malfunction
Additional Manufacturer Narrative
The clip brace was slightly bent and had broken near one end.Closing pressure and closing function were found to be within specifications.The bent brace and the traces of blood found on the clip indicate improper handling during the surgical procedure.Apparently force was applied to the clip by another object, first bending and finally breaking the brace.
Event or Problem Description
During a surgical procedure, undue force seems to have been applied to the clip with a foreign object, leading to breakage.The patient did not suffer any harm.No further information has been provided by the distributor, who first indicated that the malfunction occurred during use of the device, and then that it broke during cleaning and sterilization prior to first use.The second statement does not hold, since the returned device showed traces of blood.