• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDICON E.G. YASARGIL CLIP; ANEURYSM CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDICON E.G. YASARGIL CLIP; ANEURYSM CLIP Back to Search Results
Catalog Number 586008
Medical Device Problem Codes Material Twisted/Bent (2981); Device Handling Problem (3265)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The jaw of the clip was misaligned, which can only be caused by applying great external force to the jaw.Closing function was impaired.The state of the device indicates improper handling during the surgical procedure.
 
Event or Problem Description
Application of excessive force to the jaw part of the clip during a surgical procedure caused clip jaw misalignment and impaired clip function by making it impossible to close the clip properly.The patient did not suffer any harm.No further information has been provided by the distributor, who first indicated that the malfunction occurred during use of the device, and then that it was damaged during cleaning and sterilization prior to first use.The second statement does not hold, since the returned device showed traces of blood.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
YASARGIL CLIP
Common Device Name
ANEURYSM CLIP
Manufacturer (Section D)
MEDICON E.G.
gaensaecker 15
tuttlingen, 78532
GM  78532
MDR Report Key8058151
Report Number8010099-2018-00005
Device Sequence Number1910744
Product Code HCH
UDI-Device Identifier04046826094171
UDI-Public4046826094171
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K010908
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number586008
Device Lot NumberQ388286
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2017
Initial Report FDA Received Date11/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-