| Model Number |
GIA6038RL |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the functional end to end reconstruction of transverse colon in a colectomy procedure, when the surgeon tried to perform the 2nd firing, the firing stopped midway.The procedure was completed with another device.The tissue was not stiff or thick.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The loading unit was received partially applied with the interlock engaged.No damage was noted to the knife blade.The single use loading unit (sulu) was reset and loaded into a test unit.The device and loading unit were applied to the appropriate test media with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.If the firing knob is not fully retracted, difficulty in unloading the loading unit may be experienced, and may prevent further loading from occurring.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|